NanoPhotometer® NP80 Tech Specs
Microvolume and Cuvette Functionality | Starting at 0.3 µl | Scan Range 200 – 900 nm
General Specifications
Dimensions: 20 cm x 20 cm x 12 cm
Weight: 3.8 – 5.2 kg depending on configuration
Processing Power/Software
- Linux based NPOS
- Intel Celeron dual core 2.4 GHz
- 64 GB Internal Data Storage
- 2x USB A, USB B
- HDMI, Ethernet, WiFi
- Windows 8, 10 (32 & 64 bit)
- OS X (Intel x86 and Apple M1)
- iOS and Android OS
Display
- 1024 x 600 pixels
- Glove compatible touchscreen
Built-in Battery Pack
Optional rechargeable lithium ion battery:
95 Wh, 6.6 Ah, 8 h
Min. charging cycles: 800
NanoVolume
Performance
Detection Range dsDNA | 1 – 16,500 ng/µl |
Detection Range BSA | 0.03 – 478 mg/ml |
Minimum Sample Size | 0.3 – 2 µl |
Photometric Range (10 mm equivalent) | 0.02 – 330 A |
Path Length | 0.67 and 0.07 mm |
Dilution Factor | 15 and 140 |
Vortex | 2,800 rpm; tube size up to 2.0 ml |
Cuvette
Performance
Detection Range dsDNA | 0.1 – 130 ng/µl |
Detection Range BSA | 0.003 – 3.7 mg/ml |
Photometric Range | 0 – 2.6 A |
Center Height (Z-Height) | 8.5 mm |
Cell Types | Outside dimension 12.5 x 12.5 mm |
Heating | 37°C ± 0.5°C |
Optical
Specifications
Wavelength Scan Range | 200 – 900 nm |
Measure Time For Full Scan Range | 2.5 – 4 seconds |
Wavelength Reproducibility | ± 0.2 nm |
Wavelength Accuracy | ± 0.75 nm |
Bandwidth | < 1.5 nm |
Absorbance Reproducibility | (Cuvette): < 0.002 A @ 0 – 0.3 A @ 280 nm CV < 1% @ 0.3 – 2.0 A @ 280 nm (Lid 15): < 0.002 A @ 0 – 0.3 A @ 280 nm CV < 1% @ 0.3 – 1.7 A @ 280 nm |
Absorbance Accuracy | < 1.75% @ 0.7 A @ 280 nm of the reading |
Stray Light | < 0.5% @ 240 nm using NaI |
Optical Arrangement | 1 x 4096 CMOS Array |
Lamp | Xenon flash lamp |
Lifetime | 109 flashes, up to 10 years |
General
Specifications
Operating Voltage | 90 – 250 V ± 10%, 50/60 Hz, 90 W, 18/19 VDC |
Certification | CE, IEC 61010-1:2012 and EN 61326-1:2013 |
Battery Certification | IEC 62133 and UN38.3 transport test |
Security | Slot for Kensington lock |